Clinical Chemistry AACC Online Job Center
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 32: 1460-1463, 1986;
This Article
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Mullins, R. E.
Right arrow Articles by Austin, G. E.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Mullins, R. E.
Right arrow Articles by Austin, G. E.

Clinical Chemistry, Vol 32, 1460-1463, Copyright © 1986 by American Association for Clinical Chemistry

Sensitivity of isoelectric focusing, ion exchange, and affinity chromatography to labile glycated hemoglobin

RE Mullins and GE Austin

We examined the effect of labile glycated hemoglobin on measurements of glycated hemoglobin by several commercial procedures. Erythrocytes from diabetic and nondiabetic patients were incubated in vitro with various concentrations of glucose, to generate labile glycated hemoglobin, and the species of glycated and nonglycated hemoglobin in each sample were identified by isoelectric focusing. Glycated hemoglobin was then assayed by the Bio-Rad A1 column method (I), the Bio-Rad A1c column method (II), and the Pierce affinity column method (III). I was sensitive to the labile (aldimine) fraction of glycated hemoglobin, and percentages of glycated hemoglobin so determined represented the sum of the labile fraction plus hemoglobin A1c and other stable glycated species. This spurious increase in glycated hemoglobin concentration by the aldimine could be obviated by any of three wash procedures, which eliminated the labile fraction from the samples: incubating the erythrocytes (a) in phosphate-buffered saline (pH 7.4) for 18 h at 22 degrees C, (b) in 100 mmol/L citrate (pH 5.0) for 30 min at 37 degrees C, or (c) in saline containing, per liter, 30 mmol of semicarbazide and 12 mmol of aniline hydrochloride (pH 5.0) for 30 min at 37 degrees C. Methods II and III did not detect the labile fraction. However, treatments a-c decreased the concentrations of stable glycated hemoglobin as determined by all three column-chromatographic methods as compared with unwashed sample. By isoelectric focusing we determined that blood with high glucose content had concentrations of aldimine roughly proportional to the blood glucose concentration. The kinetics of formation of labile glycated hemoglobin in these cells were consistent with the reported rate constants determined by using purified hemoglobin preparations in vitro.


The following articles in journals at HighWire Press have cited this article:


Home page
Clin. Chem.Home page
I. Gibb, A. Parnham, M. Fonfrede, and F. Lecock
Multicenter Evaluation of Tosoh Glycohemoglobin Analyzer
Clin. Chem., October 1, 1999; 45(10): 1833 - 1841.