Clinical Chemistry Link to Randox Laboratories Web Site
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 33: 138-139, 1987;
This Article
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Dujardin, B. C.
Right arrow Articles by van der Ven-Jongekrijg, J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Dujardin, B. C.
Right arrow Articles by van der Ven-Jongekrijg, J.

Clinical Chemistry, Vol 33, 138-139, Copyright © 1987 by American Association for Clinical Chemistry

Influence of ambient temperature on the Ektachem DT60 results for total bilirubin and total protein

BC Dujardin, JC Hafkenscheid, AF Roijers, M Schoorl and J van der Ven-Jongekrijg

We evaluated the Kodak Ektachem DT60 analyzer with the DTE module in a two-month clinical trial before its introduction in the Netherlands. At ambient temperatures that differed from that at which the DT60 was calibrated, aberrant behavior for total bilirubin and total protein assays was observed. At subnormal temperatures the former gave higher results than expected, and the latter, lower results. We found a similar effect for total protein determined by the manual biuret reaction at different temperatures, when results were calculated from calibration at a fixed temperature. Total bilirubin assayed by the manual diazobilirubin method showed no such effect. Although the DT60 analyzer is equipped with a temperature-regulated incubator, we conclude that the manufactures's recommended room temperature range for these assays should be narrowed and the three-month calibration period adjusted according to external circumstances.