Clinical Chemistry
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 33: 1756-1760, 1987;
This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Turpeinen, U.
Right arrow Articles by Stenman, U. H.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Turpeinen, U.
Right arrow Articles by Stenman, U. H.

Clinical Chemistry, Vol 33, 1756-1760, Copyright © 1987 by American Association for Clinical Chemistry

Liquid-chromatographic determination of beta 2-microglobulin, alpha 1- acid glycoprotein, and albumin in urine

U Turpeinen, E Koivunen and UH Stenman
Helsinki University Central Hospital, Department I of Obstetrics and Gynecology, Finland.

We report a reversed-phase "high-performance" liquid-chromatographic assay for beta 2-microglobulin, alpha 1-acid glycoprotein, and albumin in urine. The urine samples are chromatographed on a 15 X 3.2 mm C18- column with a solvent gradient between two mixtures: 1 mL of trifluoroacetic acid per liter of water and 1 mL of trifluoroacetic acid per liter of acetonitrile. The proteins are detected at 218 nm. The useful lower limit of sensitivity for quantification is 0.1 micrograms of each protein in 100-microliters injected samples, corresponding to 1 mg of protein per liter of urine. The linear measuring range extends from 0.1 to 100 micrograms of protein for each peak. Mean analytical recoveries ranged from 92 to 100%. The mean within-assay CV was 7.5%, the between-assay CV 15%. Comparison of results with those obtained by immunoturbidimetry and radioimmunoassay shows acceptable correlation. Reproducible and rapid with low operational cost, this method is suitable for use in routine screening and quantification of urinary proteins.


The following articles in journals at HighWire Press have cited this article:


Home page
Clin. Chem.Home page
D. Sviridov, B. Meilinger, S. K. Drake, G. T. Hoehn, and G. L. Hortin
Coelution of Other Proteins with Albumin during Size-Exclusion HPLC: Implications for Analysis of Urinary Albumin
Clin. Chem., March 1, 2006; 52(3): 389 - 397.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1987 by the American Association for Clinical Chemistry.