Clinical Chemistry
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 33: 1891-1895, 1987;
This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Armitage, E. K.
Right arrow Articles by Miller, W. G.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Armitage, E. K.
Right arrow Articles by Miller, W. G.

Clinical Chemistry, Vol 33, 1891-1895, Copyright © 1987 by American Association for Clinical Chemistry

Optical kinetic method for calibration of spectrophotometer temperature: demonstration with the Cobas-Bio analyzer

EK Armitage and WG Miller
Department of Pathology, Medical College of Virginia/Virginia Commonwealth University, Richmond 23298-0597.

The pseudo-first-order rate constant for the Jaffe reaction with creatinine varies logarithmically with temperature and was calibrated in the range 25 to 37 degrees C to measure the temperature of the liquid in the lightpath of spectrophotometric instrumentation. The reagent concentrations can be adjusted to permit rate-constant measurements in time intervals from a few seconds to several minutes. The temperature increment that can be resolved is limited only by the analytical imprecision of the instrumentation used to measure the rate constant and the calibration temperature. In this investigation, a temperature SD of 0.03 degrees C could be measured. Two Cobas-Bio centrifugal analyzers, used to demonstrate the utility of this technique, were found to have temperature errors from -1.0 to -1.7 degrees C in the 30 to 37 degrees C range and overall temperature SD of 0.19 and 0.36 degree C, respectively, at 37 degrees C. Analysis of variance gave between-rotor SD of 0.14 and 0.34 degrees C and within- rotor SD of 0.13 and 0.11 degree C, respectively. We found temperature differences of 0.3 degree C between cuvets in a rotor, and gradients of 0.3 and 0.4 degree C, respectively, from the top to bottom of an individual cuvet in the two instruments.