Clinical Chemistry
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 33: 269-272, 1987;
This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Howey, J. E.
Right arrow Articles by Fraser, C. G.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Howey, J. E.
Right arrow Articles by Fraser, C. G.

Clinical Chemistry, Vol 33, 269-272, Copyright © 1987 by American Association for Clinical Chemistry

Assay of serum fructosamine that minimizes standardization and matrix problems: use to assess components of biological variation

JE Howey, MC Browning and CG Fraser

Methodological problems with the assay of fructosamine in serum-- standardization, matrix effects, and dependence on buffer pH--have been minimized with a method involving colorimetric assay of each specimen and subsequent re-assay after standard addition of 1-deoxy-1- morpholinofructose. Absorbance at optimum wavelength of 540 nm varies linearly with fructosamine concentration to at least 5.5 mmol/L, and between-run precision is about 6% for both patients' specimens and quality control materials. Correction of fructosamine to serum albumin of 40 g/L minimizes the effect of albumin while maintaining transferability of data and reference values. From data on biological variation, the analytical goal for precision (CV) is less than or equal to 2.6%. The square root of the ratio of intra- to interindividual variance is low, indicating that fructosamine concentrations have a high index of individuality; thus conventional population-based reference values are of limited use. Although this assay may be useful in monitoring disease, we doubt that it provides a valid screening test for diabetes.


The following articles in journals at HighWire Press have cited this article:


Home page
Clin. Chem.Home page
K. Fujita, L. K. Curtiss, I. Sakurabayashi, F. Kameko, N. Okumura, F. Terasawa, M. Tozuka, and T. Katsuyama
Identification and Properties of Glycated Monoclonal IgA That Affect the Fructosamine Assay
Clin. Chem., May 1, 2003; 49(5): 805 - 808.
[Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1987 by the American Association for Clinical Chemistry.