Clinical Chemistry
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 33: 563-566, 1987;
This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Klee, G. G.
Right arrow Articles by Reynoso, G.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Klee, G. G.
Right arrow Articles by Reynoso, G.

Clinical Chemistry, Vol 33, 563-566, Copyright © 1987 by American Association for Clinical Chemistry

Discrepancies in carcinoembryonic antigen measurements: survey and control values vs values for patients

GG Klee, LA Dodge and G Reynoso

We analyzed the carcinoembryonic antigen (CEA) test results reported in the College of American Pathologists' (CAP) surveys to determine the relationship between the source of CEA used to manufacture the survey specimens and the discrepancies among analytical methods. With the 1983 survey specimens, which were prepared from metastatic colon carcinoma, laboratories using Roche RIA with Clinetics columns reported results that were only one-half the values reported by laboratories using the Abbott polyclonal enzyme immunoassay. With the 1984 specimens, prepared from a different metastatic colon carcinoma, and the 1985 specimens, prepared from a tissue-culture source of CEA, the Roche results were about one-sixth as large as the Abbott results. These differences are larger than the reported assay differences for patients' specimens. In addition, twofold proportional differences were found when survey and control specimens were tested with different lots of Abbott polyclonal reagent, whereas only random differences were found with 102 patients' specimens. Evidently, assay systems perform differently with proficiency-testing and control specimens than with patients' specimens.