Clinical Chemistry
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 33: 1233-1235, 1987;
This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Forest, J. C.
Right arrow Articles by Gosselin, M.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Forest, J. C.
Right arrow Articles by Gosselin, M.

Clinical Chemistry, Vol 33, 1233-1235, Copyright © 1987 by American Association for Clinical Chemistry

Quality-control scheme for blood glucose measured outside the laboratory

JC Forest, F Rousseau, R Carrier, P Hivon and M Gosselin

We propose a quality-control (QC) scheme to evaluate the performance of blood glucose estimates made with a reflectometer in hospital wards or at home. We constructed a chart illustrating the comparison by regression analysis of the results of 254 reflectometer readings (y) and simultaneous capillary blood glucose analyses in the laboratory (x). Arbitrary acceptance limits around the regression line (y = 0.5 + 0.87x; r = 0.97, Sy/x = 0.8 mmol/L) were established, based on the mean +/- 2 SD of the reflectometer readings of a series of classes of 1.1 mmol/L interval, made with the results from the laboratory analyzer. Each ward using a reflectometer is provided with the QC chart, and the first blood glucose estimation with the reflectometer on each working shift is controlled by re-assay in the laboratory. If the reflectometer reading is outside the limits, the next glucose estimation is also re- assayed. If two consecutive readings are outside the limits, the reflectometer procedure is considered to be failing, the instrument's readings are not used for medical decisions, and remedial action is taken. The chart can also be used as a teaching tool for training the patients or the nurses and to evaluate individual performance.


The following articles in journals at HighWire Press have cited this article:


Home page
Clin. Chem.Home page
M. du Plessis, J. B. Ubbink, and W.J. H. Vermaak
Analytical Quality of Near-Patient Blood Cholesterol and Glucose Determinations
Clin. Chem., August 1, 2000; 46(8): 1085 - 1090.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1987 by the American Association for Clinical Chemistry.