Clinical Chemistry
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 34: 127-130, 1988;
This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Morisi, G.
Right arrow Articles by Menotti, A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Morisi, G.
Right arrow Articles by Menotti, A.

Clinical Chemistry, Vol 34, 127-130, Copyright © 1988 by American Association for Clinical Chemistry

Improved determination of selenium in serum by Zeeman atomic absorption spectrometry

G Morisi, M Patriarca and A Menotti
Laboratorio di Biochimice, Clinica, Institute Superiore di Sanita Roma, Italy.

We have improved Zeeman atomic absorption spectrometric determination of selenium in serum by using an acidic solution of Ag + Cu + Mg as a matrix modifier and performing the charring step in flowing O2. Under these conditions, we could calibrate serum results with selenium standards in bovine albumin solutions. Mean analytical recovery was 98%, and the CV was 1.8% within runs, 2.9% between runs. Analysis of Reference Materials from the U.S. National Bureau of Standards (SRM 909 and RM 8419) yielded the values of 106 (SD 2.4) and 15 (SD 1.6) microgram/L, respectively, in good agreement with the expected values (106 and 16, respectively). The method--being reliable and relatively simple and rapid--is suitable for use in epidemiological screenings. Mean selenium concentrations in serum sampled from 274 adults in central Italy were 90 (SD 15) microgram/L. For 11-year-old children, these values were lower and showed a tendency to sex-related difference: 82 (SD 9.9) microgram/L for 97 boys, 78 (SD 9.3) microgram/L for 90 girls.


The following articles in journals at HighWire Press have cited this article:


Home page
J Child NeurolHome page
T. A. Glauser, M. Titanic-Schefft, and C.E. Pippenger
Racial Differences in Free Radical Scavenging Enzyme Activity in Children
J Child Neurol, June 1, 1999; 14(6): 382 - 387.
[Abstract] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1988 by the American Association for Clinical Chemistry.