Clinical Chemistry
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 34: 2039-2043, 1988;
This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Ohlson, S.
Right arrow Articles by Larsson, P. O.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Ohlson, S.
Right arrow Articles by Larsson, P. O.

Clinical Chemistry, Vol 34, 2039-2043, Copyright © 1988 by American Association for Clinical Chemistry

High-performance liquid affinity chromatography: rapid immunoanalysis of transferrin in serum

S Ohlson, BM Gudmundsson, P Wikstrom and PO Larsson
Perstorp Biolytica AB, Lund, Sweden.

We describe a new method for quantitatively measuring substances of clinical interest by high-performance liquid affinity chromatography (HPLAC). As a model system we selected analysis for transferrin in human serum with immobilized antibodies in a high-performance liquid chromatographic system. SelectiSpher-10 Activated Tresyl columns (5 or 10 x 0.5 cm) were used for in situ coupling of polyclonal antibodies to transferrin. The amount of transferrin eluted was determined by integrating the eluted peak at 280 nm. The whole analytical procedure-- including injection of sample, washing, elution, and analysis of data-- takes only 7 min. We characterized the HPLAC system for analysis of transferrin in several ways: intra-assay CV approximately 3%; inter- assay CV 2-9%; linear response up to 1 mg/mL column volume; detection limit approximately 3 micrograms; analytical recovery 98% +/- 2%; purity of eluted sample greater than 95% (SDS-PAGE). The HPLAC method was compared with "rocket" immunoelectrophoresis, a commonly used method of analysis for transferrin, and there was excellent correlation between the two methods (r = 0.96, n = 60). Benefits of this HPLAC technique include high precision, rapid analysis, and simplified sample handling.


The following articles in journals at HighWire Press have cited this article:


Home page
Clin. Chem.Home page
D. S. Hage
Affinity Chromatography: A Review of Clinical Applications
Clin. Chem., May 1, 1999; 45(5): 593 - 615.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1988 by the American Association for Clinical Chemistry.