Clinical Chemistry
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 34: 2542-2545, 1988;
This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Sedor, F. A.
Right arrow Articles by Schneider, K. A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Sedor, F. A.
Right arrow Articles by Schneider, K. A.

Clinical Chemistry, Vol 34, 2542-2545, Copyright © 1988 by American Association for Clinical Chemistry

Reflotron cholesterol measurement evaluated as a screening technique

FA Sedor, CM Holleman, S Heyden and KA Schneider
Hospital Laboratories, Duke University Medical Center, Durham, NC 27710.

We evaluated the analytical performance of Boehringer Mannheim Diagnostics' "Reflotron" analyzer for the measurement of cholesterol. Coefficients of variation (CVs) for whole-blood cholesterol were: within-day 2.0% and 2.2% at 1680 and 2670 mg/L, respectively; between- day 1.8% and 2.4% (n = 9 and 8). Results were similar for serum and heparinized or EDTA-treated single-donor plasma (CV 1.4% to 2.6%). CVs of results for two reconstituted commercial quality-control materials were 3.4% and 4.6%. Heparin and hematocrit were evaluated as interferents, and critical limits for interference were identified for bilirubin, hemoglobin, and triglyceride in blood and plasma or serum. When sample collection and analysis were controlled by trained personnel, results with the Reflotron (y) compared well with those by the Ektachem procedure (x) for both blood and serum samples: r = 0.950, y = 0.944x + 130 mg/L; and r = 0.955, y = 0.93x + 43.5 mg/L, respectively. The same comparability was observed when the analysis was performed by briefly trained high-school students: r = 0.980, y = 0.949x + 23 mg/L. Performance decreased when both collection and analysis were performed by laymen: r = 0.880, y = 0.870x + 186 mg/L.


The following articles in journals at HighWire Press have cited this article:


Home page
JAMAHome page
J. L. Furst
Portable Cholesterol Analyzers
JAMA, September 5, 1990; 264(9): 1101 - 1101.
[Abstract] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1988 by the American Association for Clinical Chemistry.