Clinical Chemistry Link to Randox Laboratories Web Site
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 34: 746-748, 1988;
This Article
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Guagnellini, E.
Right arrow Articles by Stella, P.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Guagnellini, E.
Right arrow Articles by Stella, P.

Clinical Chemistry, Vol 34, 746-748, Copyright © 1988 by American Association for Clinical Chemistry

Reliability of IL Monarch ion-selective electrode module for sodium, potassium, and chloride measurements

E Guagnellini, G Spagliardi, G Bernardi and P Stella
Laboratorio Analisi, Ospedale Bassini, Milano, Italy.

We evaluated the IL Monarch random-access centrifugal analyzer for measurement of Na+, K+, and Cl- by an indirect potentiometric method. For different concentrations of control material, the total precision (CV) ranged between 0.82% and 1.14% for the three electrolytes; linearity was acceptable within a range of 103 to 215 mmol/L for Na+, 1.6-15.25 mmol/L for K+, and 80-173 mmol/L for Cl-. Data correlated well with those by flame photometry for Na+ and K+ and with those by coulometry for Cl-, both for various biological materials--sera, urines, dialysis fluids--and commercial control materials from various producers. Stability of the potentiometric signal was acceptable: daily variations were 0.2 mV for Na+, 0.05 mV for K+, and 0.03 mV for Cl-. Accordingly, we conclude that the system supplies reproducible and accurate results while being easy to use and requiring little maintenance. The use of indirect potentiometry offers results consistent with those obtained with traditional methods, and easily interpretable by clinical staff. However, better information about the actual ion activity in the tested sample for certain pathologies such as hyperlipemia and dysproteinemia could be obtained by methods involving direct potentiometry.