Clinical Chemistry 34: 1749-1752, 1988;
This Article
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Wilkins, T. A.
Right arrow Articles by Cambiaso, C. L.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Wilkins, T. A.
Right arrow Articles by Cambiaso, C. L.

Clinical Chemistry, Vol 34, 1749-1752, Copyright © 1988 by American Association for Clinical Chemistry

High sensitivity, homogeneous particle-based immunoassay for thyrotropin (Multipact)

TA Wilkins, G Brouwers, JC Mareschal and CL Cambiaso
Acade Diagnostic Systems SA, Louvain La Neuve, Belgium.

We describe the first homogeneous, nonradioactive, high-sensitivity assay for human thyrotropin (TSH). The assay is based on particle immunoassay techniques, wherein 800-nm particles form the basis for the immunochemistry, delivery, and the detection technologies, respectively. Our assay also is the first to involve the use of fragmented monoclonal antibodies (to eliminate serum interferences) covalently coupled to particles without loss of their binding properties. Assays are performed in a semiautomated mode with use of a new modular system (Multipact). Equilibrium is reached in less than 2 h. Precision profile, sensitivity, and clinical studies indicate that the assay is accurate, has good precision at low concentrations, and that detection-limit characteristics compare well with those of a leading commercial high-sensitivity immunoradiometric assay (IRMA) for TSH. Dilution characteristics were satisfactory down to the assay's detection limit for a range of clinical samples. Correlation studies vs a reference IRMA method yielded the regression equation, present method = 0.976 (IRMA) + 0.002 milli-int. unit/L (r = 0.98), for 223 samples with TSH concentrations in the range 0 to 30 milli-int. units/L. For 40 samples with TSH less than or equal to 1.0 milli-int. unit/L it was: present method = 0.94 (IRMA) + 0.005 milli-int. unit/L (r = 0.96).


The following articles in journals at HighWire Press have cited this article:


Home page
Clin. Chem.Home page
A. S. Negi and A. K. Sood
Electric Field-Enhanced Sensitivity of Grafted Ligands and Receptors
Clin. Chem., February 1, 2008; 54(2): 366 - 370.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1988 by the American Association for Clinical Chemistry.