Clinical Chemistry, Vol 35, 2226-2227, Copyright © 1989 by American Association for Clinical Chemistry
R Hahnel, MH Avenell and E Hahnel
Department of Obstetrics & Gynaecology, University of Western Australia, King Edward Memorial Hospital for Women, Subiaco.
Estrogen receptor (ER) and progesterone receptor (PR) assays were performed by solid-phase enzyme immunoassays (Abbott) on low-speed (2600 x g) and high-speed (100,000 x g) supernates of breast-cancer homogenates. The results were similar, indicating that ultracentrifugation is not essential for ER and PR assays with the Abbott kit.