Clinical Chemistry Link to Randox Laboratories Web Site
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 35: 2236-2241, 1989;
This Article
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Bachorik, P. S.
Right arrow Articles by Cloey, T. A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Bachorik, P. S.
Right arrow Articles by Cloey, T. A.

Clinical Chemistry, Vol 35, 2236-2241, Copyright © 1989 by American Association for Clinical Chemistry

Rate immunonephelometry and radial immunodiffusion compared for apolipoproteins AI and B assay

PS Bachorik and TA Cloey
Department of Pediatrics, Johns Hopkins University School of Medicine,Baltimore, MD 21205.

We measured apolipoproteins (apo) AI and B in fresh plasma samples and serum control pools by using two rate immunonephelometric assay (INA) methods (Beckman "Array" and Behring) and radial immunodiffusion (RID). Both INA methods on average gave apoAI values similar to those by RID in fresh plasma samples. The coefficients of correlation for the two INA-RID method pairs were as follows: Beckman INA vs RID, n = 94, r = 0.73; and Behring INA vs RID, n = 112, r = 0.66. The bias between the INA and RID methods was reflected reasonably well by either lyophilized or frozen control pools. For apoB, the Beckman INA measurements in fresh plasma averaged 40% lower than RID values, due almost entirely to the values assigned to calibration sera, but results by the two methods were highly correlated (n = 94, r = 0.93). The Behring INA values for fresh plasma averaged only 13% lower than RID values, but the two methods were less well correlated (n = 112, r = 0.74). Frozen control pools were most suitable for apoB analysis.