Clinical Chemistry AACC Online Job Center
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 35: 461-463, 1989;
This Article
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Grutzmeier, S.
Right arrow Articles by von Schenck, H.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Grutzmeier, S.
Right arrow Articles by von Schenck, H.

Clinical Chemistry, Vol 35, 461-463, Copyright © 1989 by American Association for Clinical Chemistry

Four immunochemical methods for measuring C-reactive protein in plasma compared

S Grutzmeier and H von Schenck
Department of Internal Medicine, University Hospital, Linkoping, Sweden.

Four immunochemical methods for measuring C-reactive protein (CRP) in plasma were compared: radial immunodiffusion (RID), electroimmunoassay (EIA), immunoturbidimetry (IT), and laser nephelometry (LN). Close agreement was found between RID and EIA (EIA = 0.96 RID + 7.7 mg/L, r = 0.977, n = 100), IT and EIA (EIA = 1.11 IT - 15.8 mg/L, r = 0.951, n = 100), IT and RID (RID = 1.10 IT - 11.5 mg/L r = 0.959, n = 94). In initial studies in 1983, LN showed good agreement with RID (LN = 0.98 RID + 11.9 mg/L, r = 0.973, n = 60). Since then, normal CRP values measured by LN have tended to increase and, at the time of the present study (1986), LN agreed poorly with other methods. The reason for this change is obscure. Having excluded the possibility of enzymatic degradation of CRP, we conclude that instrumentation problems are involved, and therefore we no longer use this method.


The following articles in journals at HighWire Press have cited this article:


Home page
Am. J. Clin. Nutr.Home page
N. Pascal, E. K. S. Amouzou, A. Sanni, F. Namour, I. Abdelmouttaleb, M. Vidailhet, and J.-L. Gueant
Serum concentrations of sex hormone binding globulin are elevated in kwashiorkor and anorexia nervosa but not in marasmus
Am. J. Clinical Nutrition, July 1, 2002; 76(1): 239 - 244.
[Abstract] [Full Text] [PDF]


Home page
Clin. Chem.Home page
E. M. Macy, T. E. Hayes, and R. P. Tracy
Variability in the measurement of C-reactive protein in healthy subjects: implications for reference intervals and epidemiological applications
Clin. Chem., January 1, 1997; 43(1): 52 - 58.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1989 by the American Association for Clinical Chemistry.