Clinical Chemistry
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 35: 1362-1366, 1989;
This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Klee, G. G.
Right arrow Articles by Post, G.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Klee, G. G.
Right arrow Articles by Post, G.

Clinical Chemistry, Vol 35, 1362-1366, Copyright © 1989 by American Association for Clinical Chemistry

Effect of counting errors on immunoassay precision

GG Klee and G Post
Department of Laboratory Medicine and Pathology, Mayo Foundation, Rochester 55905.

Using mathematical analysis and computer simulation, we studied the effect of gamma scintillation counting error on two radioimmunoassays (RIAs) and an immunoradiometric assay (IRMA). To analyze the propagation of the counting errors into the estimation of analyte concentration, we empirically derived parameters for logit-log data- reduction models for assays of digoxin and triiodothyronine (RIAs) and ferritin (IRMA). The component of the analytical error attributable to counting variability, when expressed as a CV of the analyte concentration, decreased approximately linearly with the inverse of the square root of the maximum counts bound. Larger counting-error CVs were found at lower concentrations for both RIAs and the IRMA. Substantially smaller CVs for overall assay were found when the maximum counts bound progressively increased from 500 to 10,000 counts, but further increases in maximum bound counts resulted in little decrease in overall assay CV except when very low concentrations of analyte were being measured. Therefore, RIA and IRMA systems based in duplicate determinations having at least 10,000 maximum counts bound should have adequate precision, except possibly at very low concentrations.