Clinical Chemistry, Vol 36, 1941-1944, Copyright © 1990 by American Association for Clinical Chemistry
Y Xu, B Halsall and WR Heineman
Department of Chemistry, University of Cincinnati, OH 45221-0172.
A sandwich-type heterogeneous enzyme immunoassay with flow-injection analysis for alpha-fetoprotein (AFP) in human serum has been developed. 4-Aminophenol, the product of enzymatic reaction, is detected amperometrically. The interassay CV for this electrochemical enzyme immunoassay was less than 8.2%, with a minimum detection limit for AFP of 0.163 micrograms/L. The calibration curve had a linear range of 0.316-100 micrograms/L. Studies with 48 human maternal serum samples, comparing results by this method with those by a commercial kit, showed a good correlation (r = 0.961). This procedure provides an alternative method for determining low concentrations of AFP in human maternal serum.
The following articles in journals at HighWire Press have cited this article:
![]() |
S.-F. Chen, Y. Xu, and M. P.-C. Ip Electrochemical Enzyme Immunoassay for Serum Prostate-Specific Antigen at Low Concentrations Clin. Chem., August 1, 1997; 43(8): 1459 - 1461. [Full Text] |
||||