Clinical Chemistry
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 36: 325-329, 1990;
This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Shephard, M. D.
Right arrow Articles by Whiting, M. J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Shephard, M. D.
Right arrow Articles by Whiting, M. J.

Clinical Chemistry, Vol 36, 325-329, Copyright © 1990 by American Association for Clinical Chemistry

Falsely low estimation of triglycerides in lipemic plasma by the enzymatic triglyceride method with modified Trinder's chromogen

MD Shephard and MJ Whiting
Department of Biochemistry and Chemical Pathology, Flinders Medical Centre, Bedford Park, South Australia.

The enzymatic assay of triglyceride, based on the use of L-glycerol-3- phosphate oxidase (EC 1.1.3.21) and a modified Trinder's chromogen involving 4-chlorophenol, is subject to strong negative interference at concentrations of triglyceride greater than 20 mmol/L, such as occur in grossly lipemic plasma. This interference is caused by the rapid utilization of oxygen, resulting in the reaction becoming transiently anaerobic. The dye product already formed may then be reduced ("bleached") by acting as an alternative electron acceptor for glycerol- 3-phosphate oxidase. Reduction of the dye leads to a marked decrease in final absorbance at 505 nm. Grossly underestimated values for triglyceride concentrations, apparently within the linear range of the assay, may therefore be inadvertently obtained with equilibrium methods. We suggest that samples giving unexpectedly low results for lipemic plasma should be re-assayed after dilution or with use of a smaller volume of sample.