Clinical Chemistry, Vol 36, 723-727, Copyright © 1990 by American Association for Clinical Chemistry
DR Abendschein, HL Fontanet and R Nohara
Cardiovascular Division, Washington University School of Medicine, St. Louis, MO 63110.
We report a convenient chromatofocusing procedure for rapid and sensitive quantification of isoforms of the MM isoenzyme of creatine kinase (EC 2.7.3.2) in plasma and efficient methods for preserving isoform profiles during handling of specimens. The assay involves use of prepacked, re-usable Mono P chromatofocusing columns and a "Fast Protein Liquid Chromatograph" (FPLC) system with on-line detection of isoform enzymatic activity in column effluent. Profiles of isoforms are analyzed within 25 min with the use of a 1-mL column; the lower limit of sensitivity for CK activity is 5 mU, and recovery of each isoform is within 1% of the amount added to plasma. Collection of blood specimens in Vacutainer Tubes containing 28.5 mumol of EDTA (final concentration in plasma, 7 to 10 mmol/L) inhibited carboxypeptidase activity in plasma by 76%, sufficient to essentially abolish isoform conversion in vitro at room temperature. These methods should facilitate applications of isoform analysis for diagnosis of myocardial infarction and coronary artery recanalization.
The following articles in journals at HighWire Press have cited this article:
![]() |
T. Laperche, P. G. Steg, M. Dehoux, J. Benessiano, G. Grollier, E. Aliot, J.-M. Mossard, P. Aubry, D. Coisne, M. Hanssen, et al. A Study of Biochemical Markers of Reperfusion Early After Thrombolysis for Acute Myocardial Infarction Circulation, October 15, 1995; 92(8): 2079 - 2086. [Abstract] [Full Text] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |