Clinical Chemistry, Vol 36, 855-859, Copyright © 1990 by American Association for Clinical Chemistry
JE Tattersall, A Dawnay, C McLean and WR Cattell
Department of Nephrology, St. Bartholomew's Hospital, London, U.K.
We have developed and validated a two-site liquid-phase immunoradiometric assay (IRMA) of atrial natriuretic peptide (ANP) in unextracted human plasma. Both radiolabeled rabbit anti-ANP IgG and polyclonal mouse anti-ANP must bind to ANP for detection, and the assay is specific for peptides with both an intact C-terminus and a disulfide bridge. The assay sensitivity (detection limit) is 0.96 pmol/L, and the working range is 2.3-300 pmol/L, with the hook effect occurring above 500 pmol/L. Results for diluted plasma from normal subjects and from patients with renal failure paralleled the standard curve; analytical recovery of ANP added to such samples averaged 94%. The between- and within-assay CVs at 8 pmol/L were 10% and 5%, respectively. The assay is sufficiently sensitive and precise to detect the postural change in ANP concentrations in normal subjects.
The following articles in journals at HighWire Press have cited this article:
![]() |
A. Clerico, S. Del Ry, and D. Giannessi Measurement of Cardiac Natriuretic Hormones (Atrial Natriuretic Peptide, Brain Natriuretic Peptide, and Related Peptides) in Clinical Practice: The Need for a New Generation of Immunoassay Methods Clin. Chem., October 1, 2000; 46(10): 1529 - 1534. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |