Clinical Chemistry AACC Online Job Center
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 36: 855-859, 1990;
This Article
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Tattersall, J. E.
Right arrow Articles by Cattell, W. R.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Tattersall, J. E.
Right arrow Articles by Cattell, W. R.

Clinical Chemistry, Vol 36, 855-859, Copyright © 1990 by American Association for Clinical Chemistry

Immunoradiometric assay of atrial natriuretic peptide in unextracted plasma

JE Tattersall, A Dawnay, C McLean and WR Cattell
Department of Nephrology, St. Bartholomew's Hospital, London, U.K.

We have developed and validated a two-site liquid-phase immunoradiometric assay (IRMA) of atrial natriuretic peptide (ANP) in unextracted human plasma. Both radiolabeled rabbit anti-ANP IgG and polyclonal mouse anti-ANP must bind to ANP for detection, and the assay is specific for peptides with both an intact C-terminus and a disulfide bridge. The assay sensitivity (detection limit) is 0.96 pmol/L, and the working range is 2.3-300 pmol/L, with the hook effect occurring above 500 pmol/L. Results for diluted plasma from normal subjects and from patients with renal failure paralleled the standard curve; analytical recovery of ANP added to such samples averaged 94%. The between- and within-assay CVs at 8 pmol/L were 10% and 5%, respectively. The assay is sufficiently sensitive and precise to detect the postural change in ANP concentrations in normal subjects.


The following articles in journals at HighWire Press have cited this article:


Home page
Clin. Chem.Home page
A. Clerico, S. Del Ry, and D. Giannessi
Measurement of Cardiac Natriuretic Hormones (Atrial Natriuretic Peptide, Brain Natriuretic Peptide, and Related Peptides) in Clinical Practice: The Need for a New Generation of Immunoassay Methods
Clin. Chem., October 1, 2000; 46(10): 1529 - 1534.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1990 by the American Association for Clinical Chemistry.