Clinical Chemistry
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 36: 865-869, 1990;
This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by May, W.
Right arrow Articles by Maberly, G.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by May, W.
Right arrow Articles by Maberly, G.

Clinical Chemistry, Vol 36, 865-869, Copyright © 1990 by American Association for Clinical Chemistry

Evaluation of automated urinary iodine methods: problems of interfering substances identified

W May, D Wu, C Eastman, P Bourdoux and G Maberly
Australian Centre for the Control of Iodine Disorders, Westmead Hospital, Sydney, Australia.

We evaluated automated methods for measurement of urinary iodine (UI) over a range expected in iodine-replete and iodine-deficient populations. Results obtained with Technicon AutoAnalyzer II systems, based on either dialysis or acid digestion, were compared with those obtained by a manual alkaline ashing technique. Results of automated dialysis were consistently higher than those obtained by the other methods. The apparently higher concentrations of UI we measured were due to interfering substances crossing the dialysis membrane and participating in the catalytic reaction. Thiocyanate (SCN) was one endogenous substance contributing to the increased measurement of UI. For urinary SCN concentrations of 5 to 15 mg/L, the amount of overestimation in the UI measurement attributable to SCN ranged from 21.8 to 61 micrograms/L. However, SCN may account for only 40-50% of the apparent increase in UI. In samples with lower UI (less than 50 micrograms/L), interfering substances produced a 100% error in results. We conclude that the automated dialysis system should not be used to assess iodine-deficient populations. This leaves a major dilemma for researchers wanting to assess the iodine status of populations, because the automated digestion method is no longer commercially available.