Clinical Chemistry
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 36: 1502-1505, 1990;
This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Jorgensen, P. J.
Right arrow Articles by Botker, H. E.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Jorgensen, P. J.
Right arrow Articles by Botker, H. E.

Clinical Chemistry, Vol 36, 1502-1505, Copyright © 1990 by American Association for Clinical Chemistry

Analytical evaluation of a sensitive enzyme immunoassay for determinations of creatine kinase isoenzyme MB

PJ Jorgensen, M Horder, J Selmer and HE Botker
Department of Clinical Chemistry, Odense University Hospital, Denmark.

We have evaluated a new sensitive immunometric assay for the determination of creatine kinase (CK; EC 2.7.3.2) MB isoenzyme (NovoClone CK-MB), involving an enzyme label and two monoclonal antibodies directed against the B subunit and the M subunit, respectively. The anti-CK-B antibodies are bound to the solid phase. The assay was modified to be extremely sensitive and thus to measure the concentration range below and close to the cutoff value used for the diagnosis of myocardial infarction. A reference interval of 0-6 micrograms/L was found for 315 outpatients without myocardial diseases (132 men and 183 women); the overall median of the log-gaussian distribution was 1.91 micrograms/L (2.03 and 1.79 micrograms/L for men and women, respectively). Total and within-assay imprecision (CV) was less than 6% at the upper reference limit. The detection limit was 0.1 microgram/L. The assay provides a favorable signal-to-noise ratio: the calibrators 0.0, 2.0, and 30.0 micrograms/L give absorbances at 492 nm of 0.040, 0.140, and 1.600 A. respectively. We conclude that the assay provides biochemical identification of individuals with myocardial damage but without myocardial infarction.


The following articles in journals at HighWire Press have cited this article:


Home page
Clin. Chem.Home page
J. Ellenius, T. Groth, B. Lindahl, and L. Wallentin
Early assessment of patients with suspected acute myocardial infarction by biochemical monitoring and neural network analysis
Clin. Chem., October 1, 1997; 43(10): 1919 - 1925.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1990 by the American Association for Clinical Chemistry.