Clinical Chemistry Link to Randox Laboratories Web Site
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 36: 1629-1632, 1990;
This Article
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Westgard, J. O.
Right arrow Articles by Burnett, R. W.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Westgard, J. O.
Right arrow Articles by Burnett, R. W.

Clinical Chemistry, Vol 36, 1629-1632, Copyright © 1990 by American Association for Clinical Chemistry

Precision requirements for cost-effective operation of analytical processes

JO Westgard and RW Burnett
Department of Pathology, University of Wisconsin, Madison 53792.

Current performance criteria for analytical methods are often based on recommendations developed many years ago. A common criterion for imprecision requires that two times the standard deviation (s) of the method be less than the allowable total error (TEa). Unfortunately, when this criterion is minimally satisfied, commonly used control procedures cannot achieve reliable detection of medically important errors. Studies of the power functions for statistical quality-control (QC) procedures show that the magnitude of medically important errors must be at least 3s to fall near the plateau of the power curves and be readily detected by current QC procedures. For methods that just meet the precision criterion 2s less than TEa, however, medically important errors will fall on the rising portion of the power curves and typically be detected less than half of the time. From a "reverse engineering" perspective, the 2s less than TEa criterion is inadequate because it does not allow for the known performance limitations (lack of sensitivity) of commonly used control procedures. A strong case can be made for using a criterion of at least '4s less than TEa, which calls for a twofold improvement in imprecision over, the current minimum requirements. This recommendation is consistent with current industrial guidelines for process capability and would lead to more reliable detection of medically important errors.


The following articles in journals at HighWire Press have cited this article:


Home page
Clin. Chem.Home page
G. S. Cembrowski and R. N. Carey
Adding Value to Proficiency Testing Programs
Clin. Chem., January 1, 2000; 46(1): 7 - 8.
[Full Text] [PDF]


Home page