Clinical Chemistry AACC Online Job Center
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 37: 94-98, 1991;
This Article
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Ray, J. E.
Right arrow Articles by Howrie, D. L.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Ray, J. E.
Right arrow Articles by Howrie, D. L.

Clinical Chemistry, Vol 37, 94-98, Copyright © 1991 by American Association for Clinical Chemistry

Digoxin-like immunoreactivity in serum from neonates and infants reduced by centrifugal ultrafiltration and fluorescence polarization immunoassay

JE Ray, D Crisan and DL Howrie
Department of Clinical Pharmacology & Toxicology, St. Vincent's Hospital, Sydney, Australia.

We evaluated the TDx Digoxin II (Abbott) modified procedure for interference from digoxin-like immunoreactive factors (DLIF) in pediatric patients. The effectiveness of centrifugal ultrafiltration as a means of removing DLIF interference from the serum of such patients was assessed. We used sera from 40 patients who had not received digoxin, whom we divided into two age groups: 30 neonates (less than 34 days postpartum) and 10 infants (younger than six months). Digoxin-like immunoreactivity was detected in 34 of 41 (83%) neonatal specimens (range 0.2-1.0 micrograms/L) and 16 of 25 (60%) infants' specimens (range 0.2-1.3 micrograms/L). Centrifugal ultrafiltration of serum specimens from these patients reduced but did not eliminate the DLIF interference in some specimens. A comparison of concentrations of DLIF in serum with various other patients' characteristics demonstrated a strong correlation (r = 0.915; P = 0.0001) between DLIF and serum bilirubin in the infants. Apparent digoxin concentrations from 19 serum and serum ultrafiltrate samples collected from 13 patients (four neonates and nine infants) who were treated with digoxin showed a good correlation (r = 0.97); however, the serum samples showed a positive bias of 0.39 microgram/L. We conclude that the TDx Digoxin II modified procedure is still subject to considerable DLIF interference in these two pediatric populations. This interference can be reduced in some serum specimens, but cannot be eliminated completely as others reported.


The following articles in journals at HighWire Press have cited this article:


Home page
Clin. Chem.Home page
I. T. Ocal and T. R. Green
Serum digoxin in the presence of Digibind: determination of digoxin by the Abbott AxSYM and Baxter Stratus II immunoassays by direct analysis without pretreatment of serum samples
Clin. Chem., September 1, 1998; 44(9): 1947 - 1950.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1991 by the American Association for Clinical Chemistry.