Clinical Chemistry 37: 285-290, 1991;
This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Reed, P.
Right arrow Articles by McMurray, J. R.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Reed, P.
Right arrow Articles by McMurray, J. R.

Clinical Chemistry, Vol 37, 285-290, Copyright © 1991 by American Association for Clinical Chemistry

Simple, optimized liquid-chromatographic method for measuring total hydroxyproline in urine evaluated

P Reed, IB Holbrook, ML Gardner and JR McMurray
Department of Clinical Biochemistry, Hope Hospital, Salford, U.K.

We have optimized a method for measuring total hydroxyproline (HYP) in urine by HPLC after release from urinary peptides by solid-phase hydrolysis on Dowex 50W x 8 ion-exchange resin. The HYP was derivatized with 4-chlor-7-nitrobenzo-2-oxa-1,3-diazole, and excess reagent was removed with the use of a 100-mg C18 Bond-Elut cartridge. The HYP derivative was separated isocratically at 30 degrees C on a 250 x 4.6 mm reversed-phase column containing 5-microns particles of Spherisorb S5 ODS-2, with S-carboxymethylcysteine as internal standard. Total assay time was 14 min. The standard curve for the method was linear from the detection limit for HYP, 3.6 mumol/L, to 10 mmol/L. The between-batch CV was less than 5.1% and the mean analytical recovery of HYP was 95% +/- 1.4%. Comparison with a commercially available colorimetric method showed good correlation: y = 1.158x + 19.76 mumol/L (Syx = 74, n = 120), but HPLC results were 15% higher, probably from incomplete hydrolysis with the colorimetric method. This method offers a considerable improvement in assay time, specificity, sensitivity, precision, and cost compared with the colorimetric method.