Clinical Chemistry Link to Randox Laboratories Web Site
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 37: 1161-1165, 1991;
This Article
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Eckfeldt, J. H.
Right arrow Articles by Frantz, I. D.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Eckfeldt, J. H.
Right arrow Articles by Frantz, I. D., Jr

Clinical Chemistry, Vol 37, 1161-1165, Copyright © 1991 by American Association for Clinical Chemistry

Determination of serum cholesterol by isotope dilution mass spectrometry with a benchtop capillary gas chromatograph/mass spectrometer: comparison with the National Reference System's Definitive and Reference Methods

JH Eckfeldt, LA Lewis, JD Belcher, J Singh and ID Frantz Jr
Department of Laboratory Medicine and Pathology Medical School, University of Minnesota, Minneapolis 55455.

We developed an isotope dilution mass spectrometric cholesterol method with [25,26,27-13C]cholesterol as internal standard and a benchtop gas chromatograph/mass spectrometer (GC/MS) that is much easier and less time consuming than previously described Reference and Definitive Methods for cholesterol. The internal standard, cholesterol standards, and unknown specimen are delivered volumetrically with an automated dilutor and the saponifying reagent. After saponification, extraction, and derivatization, specimens are injected into a benchtop quadrupole MS with an autosampler. Unknown cholesterol concentrations are calculated automatically by comparing the peak area ratio of the m/z = 368, 371 ion pair with the ratios for the cholesterol standards (0 to 12.93 mmol/L). We found within-run and day-to-day (overall) imprecision of 0.44% and 0.95%, respectively, when specimens were assayed singly. In several lyophilized and frozen Standard Reference Material (SRM) pools, cholesterol results with our GC/MS method averaged 0.4% less than the National Institute for Standards and Technology definitive GC/MS result performed about three years earlier. Our GC/MS results averaged 1.3% and 2.0% less than results by the National Reference System (NRS) Abell-Levy-Brodie-Kendall (ALBK) results from clinical specimens and the SRM pools, respectively. These results are consistent with the previously reported bias between the NRS Reference and Definitive Methods and the 0.1% per year decrease in cholesterol concentrations in SRM pools as determined by GC/MS analysis. These results further emphasize the small but consistent bias between cholesterol results by isotope dilution mass spectrometry and the ALBK Reference Method, the latter being the basis for the National Cholesterol Education Program guidelines and population reference values.