Clinical Chemistry
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 37: 1350-1355, 1991;
This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Seiler, K.
Right arrow Articles by Simon, W.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Seiler, K.
Right arrow Articles by Simon, W.

Clinical Chemistry, Vol 37, 1350-1355, Copyright © 1991 by American Association for Clinical Chemistry

Characterization of sodium-selective optode membranes based on neutral ionophores and assay of sodium in plasma

K Seiler, K Wang, E Bakker, WE Morf, B Rusterholz, UE Spichiger and W Simon
Swiss Federal Institute of Technology (ETH), Department of Organic Chemistry, Zurich, Switzerland.

Plasticized poly(vinyl chloride) membranes that incorporate a highly lipophilic sodium-selective neutral ionophore (ETH 4120) and novel H(+)- selective chromo-ionophores, lipophilic isologs of Nile Blue, are used in the competitive binding of Na+ and H+ for a reversible, optical determination of sodium activities in buffered solutions at different selected pH values. These optode membranes are used to assay total sodium concentrations in pH-buffered human blood plasma.





HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1991 by the American Association for Clinical Chemistry.