Clinical Chemistry
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 38: 2197-2202, 1992;
This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Kershisnik, M. M.
Right arrow Articles by Ashwood, E. R.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Kershisnik, M. M.
Right arrow Articles by Ashwood, E. R.

Clinical Chemistry, Vol 38, 2197-2202, Copyright © 1992 by American Association for Clinical Chemistry

Using 16O35Cl to correct for chloride interference improves accuracy of urine arsenic determinations by inductively coupled plasma mass spectrometry

MM Kershisnik, R Kalamegham, KO Ash, DE Nixon and ER Ashwood
Department of Pathology, University of Utah Medical Center, Salt Lake City 84132.

We have observed inaccurate urine arsenic values with the method of isobaric fractionation, which was designed to correct for the 40Ar35Cl interference with 75As quantitation by inductively coupled plasma mass spectrometry. Isobaric fractionation, which is based on ion intensities at m/z 77 and 82, consistently underestimates the 40Ar35Cl interference and overestimates urine arsenic. We present an improved method for identifying the argon-chloride interference. We observed that signal intensities for the species 16O35Cl and 40Ar35Cl are proportional (I75 = 0.0295 x I51 - 14.7, r2 = 0.998; where Ix is the normalized ion intensity at m/z X) in water and urine, over a broad range of chloride concentrations (0-800 mmol/L). The proportionality constant is remarkably stable within a run (mean and SD, 0.0295 +/- 0.0023, based on 10 replicates of five chloride calibrators, 0, 100, 200, 400, and 800 mmol/L). Increased sensitivity (50-fold) for detecting the 40Ar35Cl interference provides improved accuracy for urine arsenic quantitation as demonstrated by a split-sample comparison with graphite-furnace atomic absorption spectrophotometry.


The following articles in journals at HighWire Press have cited this article:


Home page
Mineral MagHome page
D. A. Polya, P. R. Lythgoe, F. Abou-Shakra, A. G. Gault, J. R. Brydie, J. G. Webster, K. L. Brown, M. K. Nimfopoulos, and K. M. Michailidis
IC-ICP-MS and IC-ICP-HEX-MS determination of arsenic speciation in surface and groundwaters: preservation and analytical issues
Mineralogical Magazine, April 1, 2003; 67(2): 247 - 261.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1992 by the American Association for Clinical Chemistry.