Clinical Chemistry Link to Randox Laboratories Web Site
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 38: 2300-2306, 1992;
This Article
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by McBride, J. H.
Right arrow Articles by Ota, M. K.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by McBride, J. H.
Right arrow Articles by Ota, M. K.

Clinical Chemistry, Vol 38, 2300-2306, Copyright © 1992 by American Association for Clinical Chemistry

Measurement of cyclosporine by liquid chromatography and three immunoassays in blood from liver, cardiac, and renal transplant recipients

JH McBride, SS Kim, DO Rodgerson, AF Reyes and MK Ota
Department of Pathology, UCLA School of Medicine 90024-1732.

In an effort to replace HPLC for whole-blood determination of cyclosporine (CsA), we compared HPLC with radioimmunoassay (RIA; INCSTAR, Cyclo-Trac SP assay), fluorescence polarization immunoassay (FPIA; Abbott TDx), and in-house modified enzyme-multiplied immunoassay technique (EMIT; Syva Co.). For blood samples from 200 various transplant recipients, RIA = 1.262 (HPLC) - 8.16, r = 0.983; FPIA = 1.200 (HPLC) + 19.90, r = 0.981; and EMIT = 1.038 (HPLC) + 11.28, r = 0.985. For segregation by transplant type, RIA, FPIA, and EMIT demonstrated positive biases of 27%, 12%, and 3%, respectively, for liver transplant recipients (n = 50) when compared with HPLC. Heart transplant recipients (n = 50) gave positive bias values of 23%, 14%, and 4% for RIA, FPIA, and EMIT, respectively. Adult renal transplant recipients (n = 50) demonstrated positive bias values of 30%, 31%, and 0% for RIA, FPIA, and EMIT, respectively. For pediatric renal transplant recipients (n = 50), positive biases of 40%, 31%, and 9% were obtained for RIA, FPIA, and EMIT, respectively. We conclude that the modified EMIT represents the best replacement for HPLC.


The following articles in journals at HighWire Press have cited this article:


Home page
Clin. Chem.Home page
W. Steimer
Performance and Specificity of Monoclonal Immunoassays for Cyclosporine Monitoring: How Specific Is Specific?
Clin. Chem., March 1, 1999; 45(3): 371 - 381.
[Abstract] [Full Text] [PDF]


Home page
Clin. Chem.Home page
E. Schutz, D. Svinarov, M. Shipkova, P.-D. Niedmann, V. W. Armstrong, E. Wieland, and M. Oellerich
Cyclosporin whole blood immunoassays (AxSYM, CEDIA, and Emit): a critical overview of performance characteristics and comparison with HPLC
Clin. Chem., October 1, 1998; 44(10): 2158 - 2164.