Clinical Chemistry, Vol 38, 575-577, Copyright © 1992 by American Association for Clinical Chemistry
D Yucel and K Dalva
Clinical Biochemistry Laboratory, Yuksek Ihtisas Hastanesi (High Specialization Hospital), Ankara, Turkey.
Clinical chemists frequently encounter hemolyzed samples. Our study examines the effects of hemolysis on the results of 25 common biochemical tests. We collected 60 15-mL blood samples from inpatients and outpatients and mechanically hemolyzed 10 mL of the samples in a two-step procedure. We classified serum from these samples as being nonhemolyzed, moderately hemolyzed, or severely hemolyzed and then performed 25 common biochemical tests. Statistical analysis of the results showed that hemolysis had the greatest effect on the lactate dehydrogenase, acid phosphatase, and potassium tests.
The following articles in journals at HighWire Press have cited this article:
![]() |
B. Mensel, U. Wenzel, M. Roser, J. Ludemann, and M. Nauck Considerably Reduced Centrifugation Time without Increased Hemolysis: Evaluation of the New BD Vacutainer(R) SSTTMII Advance Clin. Chem., April 1, 2007; 53(4): 794 - 795. [Full Text] [PDF] |
||||
![]() |
E. J. Giltay, J. M. Geleijnse, E. G. Schouten, M. B. Katan, and D. Kromhout High Stability of Markers of Cardiovascular Risk in Blood Samples Clin. Chem., April 1, 2003; 49(4): 652 - 655. [Full Text] [PDF] |
||||
![]() |
A. C. Y. Ali and J. A. Campbell Interference of o-Raffinose Cross-Linked Hemoglobin with Routine Hitachi 717 Assays Clin. Chem., September 1, 1997; 43(9): 1794 - 1796. [Full Text] |
||||