Clinical Chemistry AACC Online Job Center
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 38: 1327-1331, 1992;
This Article
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by van der Sluijs Veer, G.
Right arrow Articles by Hoorn, R. K.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by van der Sluijs Veer, G.
Right arrow Articles by Hoorn, R. K.

Clinical Chemistry, Vol 38, 1327-1331, Copyright © 1992 by American Association for Clinical Chemistry

Temperature effects on free-thyroxine measurements: analytical and clinical consequences

G van der Sluijs Veer, I Vermes, HA Bonte and RK Hoorn
Department of Clinical Chemistry, Medisch Spectrum Twente, Enschede, The Netherlands.

We compared the analytical and clinical performance of two free- thyroxine (FT4) assays--a solid-phase radioimmunoassay, Spectria, and a time-resolved fluoroimmunoassay, Delfia, both of them two-step methods-- with the performance of a direct radioimmunoassay, Nichols, to measure FT4 concentration in equilibrium dialysate of undiluted serum. The three assays showed comparable analytical performance. We tested clinical utility in sera from 135 healthy subjects with and without thyroxine-binding abnormalities and in 61 patients with and without thyroidal illnesses. We found significant differences for FT4 measured by different assays in sera from the same euthyroid patients. To explain the differences, we studied the influence of temperature on performance and calibration. Most important was the neglected fact that the association constant for the binding of thyroxine to thyroxine- binding globulin decreases when the temperature rises from 20 to 37 degrees C, causing a doubling of FT4. The two-step assays, if performed at room temperature without a well-defined calibration, can give misleading FT4 concentrations. This is the case when sera from patients with thyroxine-binding abnormalities are measured against kit standards, made up in normal human sera. If an assay is to reflect the in vivo FT4 concentration at body temperature in all types of samples, it should be performed at body temperature. For practical reasons 37 degrees C is recommended, and reference values should be defined at 37 degrees C. The same might be valid for other free-hormone assays.


The following articles in journals at HighWire Press have cited this article:


Home page
Clin. Chem.Home page
J. E.M. Midgley
Direct and Indirect Free Thyroxine Assay Methods: Theory and Practice
Clin. Chem., August 1, 2001; 47(8): 1353 - 1363.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1992 by the American Association for Clinical Chemistry.