Clinical Chemistry, Vol 38, 1479-1482, Copyright © 1992 by American Association for Clinical Chemistry
L Ashworth, I Gibb and KG Alberti
Department of Medicine, Medical School, University of Newcastle upon Tyne, UK.
We assessed the HemoCue system for measuring glucose in 5 microL of whole blood. A glucose dehydrogenase-based reaction is used with dried reagents contained in disposable microcuvettes, which are filled with blood by capillary action. Automated hexokinase and YSI 23AM glucose analyzer methods were used for comparison. Overall imprecision (CV) was better than 4.5%, with no significant differences in results between three different HemoCue photometers and four batches of microcuvettes. Regression slopes (+/- SE) were 0.947 (0.011) with the YSI and 0.966 (0.015) with the hexokinase method. Analytical recovery of added glucose was 101-106%, and the system functioned with hematocrits up to 0.65. Bilirubin up to 453 mumol/L did not interfere, but high concentrations of endogenous (greater than 3 mmol/L) and exogenous triglycerides gave positive interference. The system proved stable and robust under a wide range of storage and handling conditions; performance was impaired only at high ambient temperature (37 degrees C). We conclude that the HemoCue system should prove useful for glucose measurement; further testing outside the laboratory is warranted.
The following articles in journals at HighWire Press have cited this article:
![]() |
K. Dungan, J. Chapman, S. S. Braithwaite, and J. Buse Glucose Measurement: Confounding Issues in Setting Targets for Inpatient Management Diabetes Care, February 1, 2007; 30(2): 403 - 409. [Full Text] [PDF] |
||||
![]() |
M. J. Pugia, G. Blankenstein, R.-P. Peters, J. A. Profitt, K. Kadel, T. Willms, R. Sommer, H. H. Kuo, and L. S. Schulman Microfluidic Tool Box as Technology Platform for Hand-Held Diagnostics Clin. Chem., October 1, 2005; 51(10): 1923 - 1932. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. B.B. Kristensen, N. G. Christensen, G. Thue, and S. Sandberg Between-Lot Variation in External Quality Assessment of Glucose: Clinical Importance and Effect on Participant Performance Evaluation Clin. Chem., September 1, 2005; 51(9): 1632 - 1636. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. D M Stork, H. Kemperman, D W. Erkelens, and T. F Veneman Comparison of the accuracy of the hemocue glucose analyzer with the Yellow Springs Instrument glucose oxidase analyzer, particularly in hypoglycemia Eur. J. Endocrinol., August 1, 2005; 153(2): 275 - 281. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. M.W. Spijkerman, M. C. Adriaanse, J. M. Dekker, G. Nijpels, C. D.A. Stehouwer, L. M. Bouter, and R. J. Heine Diabetic Patients Detected by Population-Based Stepwise Screening Already Have a Diabetic Cardiovascular Risk Profile Diabetes Care, October 1, 2002; 25(10): 1784 - 1789. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. L. M. Lynch and M. J. O'Kane Methemoglobin Interferes with the HemoCue B-Glucose Analyzer Clin. Chem., March 1, 2002; 48(3): 581 - 583. [Full Text] [PDF] |
||||
![]() |
B. Bülow, L. Hagmar, J. Eskilsson, and E. M. Erfurth Hypopituitary Females Have a High Incidence of Cardiovascular Morbidity and an Increased Prevalence of Cardiovascular Risk Factors J. Clin. Endocrinol. Metab., February 1, 2000; 85(2): 574 - 584. [Abstract] [Full Text] |
||||
![]() |
F. Y. Teng, O. Punla, and B. J. Koos Predictive Value of HemoCue Capillary Whole Blood Glucose Measurements in Pregnancy Reproductive Sciences, July 1, 1995; 2(4): 618 - 622. [Abstract] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |