Clinical Chemistry, Vol 38, 1488-1490, Copyright © 1992 by American Association for Clinical Chemistry
JC Philcox, MR Haywood and AM Rofe
Division of Clinical Chemistry, Institute of Medical and Veterinary Science, Adelaide, South Australia.
The Pharmacia Mono S HR 5/5 column has been optimized for hemoglobin A1c analysis by HPLC by using a much smaller column load, decreased buffer flow rate, and a steeper gradient than was used in previously described methods. Superior chromatographic separation, shorter analysis time, and a greatly extended column life have resulted.
The following articles in journals at HighWire Press have cited this article:
![]() |
Low HbA1c levels in a poorly controlled diabetic Postgrad. Med. J., July 1, 2003; 79(933): 421 - 421. [Full Text] [PDF] |
||||
![]() |
G. Phillipov and P. J. Phillips Components of Total Measurement Error for Hemoglobin A1c Determination Clin. Chem., October 1, 2001; 47(10): 1851 - 1853. [Full Text] [PDF] |
||||
![]() |
E. Gerlo and F. Gorus Calibration of ion-exchange HPLC measurements of glycohemoglobin: effect on interassay precision Clin. Chem., December 1, 1997; 43(12): 2353 - 2357. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |