Clinical Chemistry, Vol 39, 2063-2069, Copyright © 1993 by American Association for Clinical Chemistry
J Smith and G Osikowicz
Abbott Laboratories, North Chicago, IL 60064.
We describe a new clinical laboratory instrument, the Abbott AxSYM, which provides random- and continuous-access testing for immunoassays, 20 onboard reagents, primary tube sampling, and a throughput of 80 to 120 tests per hour. The AxSYM incorporates three separate analytical technologies for processing immunoassays: microparticle enzyme immunoassay, fluorescence polarization immunoassay, and a novel technology known as ion-capture immunoassay. The system incorporates both common and technology-specific subsystems controlled by a real- time software scheduling processor. Tests can be processed in one- or two-step sandwich or competitive formats, with variable pipetting steps, incubation periods, optical read formats, and wash sequences. Menu capabilities include tests for hepatitis, retrovirus, tumor markers, fertility markers, thyroid functions, and therapeutic drugs. The time to first result is approximately 15-25 min for most routine assays and < or = 15 min for stat assays (i.e., creatine kinase MB isoenzyme, human chorionic gonadotropin beta subunit, and therapeutic drugs). AxSYM assay performance for 23 assays was comparable with that of the Abbott IMx and TDx analyzers; specimen correlation data had correlation coefficients ranging from 0.97 to 0.99 and slopes ranging from 0.99 to 1.10. Within-run imprecision (CV) was 1.5% to 11.4%, with most assays (19 of 23) demonstrating CVs < or = 8.0%.
The following articles in journals at HighWire Press have cited this article:
![]() |
W. Dimech, L. Panagiotopoulos, J. Marler, N. Laven, S. Leeson, and E. M. Dax Evaluation of Three Immunoassays Used for Detection of Anti-Rubella Virus Immunoglobulin M Antibodies Clin. Vaccine Immunol., September 1, 2005; 12(9): 1104 - 1108. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. T. Maine, R. Stricker, M. Schuler, J. Spesard, S. Brojanac, B. Iriarte, K. Herwig, T. Gramins, B. Combs, J. Wise, et al. Development and Clinical Evaluation of a Recombinant-Antigen-Based Cytomegalovirus Immunoglobulin M Automated Immunoassay Using the Abbott AxSYM Analyzer J. Clin. Microbiol., April 1, 2000; 38(4): 1476 - 1481. [Abstract] [Full Text] |
||||
![]() |
H. A. Hendriks, W. Kortlandt, and W. M. Verweij Standardized Comparison of Processing Capacity and Efficiency of Five New-Generation Immunoassay Analyzers Clin. Chem., January 1, 2000; 46(1): 105 - 111. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Friedler, A. Raziel, M. Schachter, D. Strassburger, I. Bukovsky, and R. Ron-El Luteal support with micronized progesterone following in-vitro fertilization using a down-regulation protocol with gonadotrophin-releasing hormone agonist: a comparative study between vaginal and oral administration Hum. Reprod., August 1, 1999; 14(8): 1944 - 1948. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. H. Wilson, J. Yu, A. Karian, J. Kozlowski, and S. O'Reilly Development and Multisite Evaluation of an Automated Assay for B12 on the Abbott AxSYM Analyzer Clin. Chem., March 1, 1999; 45(3): 428 - 429. [Full Text] [PDF] |
||||
![]() |
D. H. Wilson, W. Groskopf, S. Hsu, D. Caplan, T. Langner, M. Baumann, D. DeManno, G. Williams, D. Payette, C. Dagel, et al. Rapid, automated assay for progesterone on the Abbott AxSYM(TM) analyzer Clin. Chem., January 1, 1998; 44(1): 86 - 91. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |