Clinical Chemistry AACC Online Job Center
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 39: 2235-2241, 1993;
This Article
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Beresini, M. H.
Right arrow Articles by Jaklitsch, A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Beresini, M. H.
Right arrow Articles by Jaklitsch, A.

Clinical Chemistry, Vol 39, 2235-2241, Copyright © 1993 by American Association for Clinical Chemistry

Evaluation of EMIT Cyclosporine Assay for use with whole blood

MH Beresini, D Davalian, S Alexander, R Toton-Quinn, B Barnett, MJ Cerelli, MW Hu, DE Berger, WP Blohm and A Jaklitsch
Syva Co., San Jose, CA 95135.

We evaluated the EMIT Cyclosporine Assay, designed for specific quantification of cyclosporine (CsA) in whole blood. The assay was run on the Roche Cobas Mira or Cobas Mira S analyzer. Analytical recovery from both normal donor whole blood and transplant patients' samples was within 17% of the standards. Measurement of diluted out-of-range samples also yielded excellent recovery (101-105%). Within-run, between- run, and total CVs were < or = 8.1%, 10.9%, and 12.1%, respectively. The detection limit was < 32 micrograms/L. The assay was linear from 0 to 500 micrograms/L. No significant cross-reactivity was observed for CsA metabolites AM1, AM19, and AM4N. Slight cross-reactivity occurred with metabolite AM9; 670 micrograms/L AM9 increased the measured CsA concentration by 49 micrograms/L. High concentrations of bilirubin, uric acid, triglycerides, and cholesterol, as well as 54 commonly coadministered drugs did not interfere with CsA quantification. Similarly, neither extreme values of hematocrit nor choice of anticoagulant affected CsA recovery. Sample extract stability was > 4 h, and assay calibration was stable for at least 2 weeks. Patients' samples analyzed by the EMIT assay showed strong correlation with both HPLC and 125I-RIA (r > 0.97). We conclude that the EMIT Cyclosporine Assay provides a convenient, accurate, and precise method for specific quantification of CsA in whole blood.