Clinical Chemistry
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 39: 297-304, 1993;
This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Miller, W. G.
Right arrow Articles by Chinchilli, V. M.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Miller, W. G.
Right arrow Articles by Chinchilli, V. M.

Clinical Chemistry, Vol 39, 297-304, Copyright © 1993 by American Association for Clinical Chemistry

Total error assessment of five methods for cholesterol screening

WG Miller, JM McKenney, MR Conner and VM Chinchilli
Department of Pathology, Medical College of Virginia, Virginia Commonwealth University, Richmond 23298-0597.

We report the accuracy, imprecision, total analytical errors, and patient misclassification errors for cholesterol measured from capillary whole blood, venous whole blood, and venous plasma samples by five devices used in public cholesterol screening environments: Reflotron, Vision, Ektachem DT-60, QuickRead, and Liposcan. None of the methods met the National Cholesterol Education Program (NCEP) performance recommendations of 3% CV with 3% bias. The Vision and Reflotron methods used with venous samples gave individual results with total errors consistent with a combined CV and bias in the 4-5% range; capillary blood samples had total errors > 5% (combined CV and bias criteria). The DT-60 performance was near the 5% total error criterion for capillary samples and was > 5% for venous samples. Misclassification of individuals into desirable or referral groups for venous samples was as great as 5.1% for the DT-60, 5.7% for the Vision, and 7.1% for the Reflotron. Misclassifications for capillary blood samples were as great as 6.7%, 18.3%, and 14.1% for DT-60, Vision, and Reflotron, respectively. The QuickRead and Liposcan results were substantially poorer than those obtained by the other methods.