Clinical Chemistry, Vol 39, 522-529, Copyright © 1993 by American Association for Clinical Chemistry
NE Morrissey, SF Quadri, R Kinders, C Brigham, S Rose and MJ Blend
Humana Hospital-Michael Reese, Division of Nuclear Medicine, Chicago, IL 60616.
The increasing use of monoclonal antibodies (MAbs) for disease diagnosis and therapy has created a class of patients at risk for systematic error in clinical testing due to interference by human anti- murine antibodies (HAMA). HAMA interference is often difficult to detect and can cause either an increase or a decrease in apparent concentrations of antigen present. We undertook a clinical study to test a HAMA-resistant enzyme immunoassay (EIA) format for carcinoembryonic antigen (CEA) determination. Using the Food and Drug Administration-approved CEA-EIA Monoclonal One-Step Assay (Abbott) with the addition of an acid/heat extraction of patients' specimens, we found that the resulting CEA values accurately reflected the patients' status. We demonstrated that the acid/heat-extracted specimens yield linear dilution curves and show analytical recoveries of added CEA in the range of 76-123% in HAMA-positive specimens and 86-103% in HAMA- negative specimens. The correlation of CEA values in extracted vs unextracted specimens from 184 patients and control subjects was 0.9963. The CEA detection limit of the assay was 1.6 micrograms/L for the extracted samples.
The following articles in journals at HighWire Press have cited this article:
![]() |
V. I. Luzzi, M. G. Scott, and A. M. Gronowski Negative Thyrotropin Assay Interference Associated with an IgG{kappa} Paraprotein Clin. Chem., April 1, 2003; 49(4): 709 - 710. [Full Text] [PDF] |
||||
![]() |
R. L. Bertholf, L. Johannsen, and G. Benrubi False Elevation of Serum CA-125 Level Caused by Human Anti-Mouse Antibodies Ann. Clin. Lab. Sci., October 1, 2002; 32(4): 414 - 418. [Abstract] [Full Text] [PDF] |
||||
![]() |
J.-C. Rymer, R. Sabatier, A. Daver, J. Bourleaud, M. Assicot, J. Bremond, J. Rapin, S. L. Salhi, B. Thirion, A. Vassault, et al. A New Approach for Clinical Biological Assay Comparison and Standardization: Application of Principal Component Analysis to a Multicenter Study of Twenty-One Carcinoembryonic Antigen Immunoassay Kits Clin. Chem., June 1, 1999; 45(6): 869 - 881. [Abstract] [Full Text] [PDF] |
||||