|
|
||||||||
Clinical Chemistry, Vol 39, 855-859, Copyright © 1993 by American Association for Clinical Chemistry
R Eugster, B Rusterholz, A Schmid, UE Spichiger and W Simon
Department of Organic Chemistry, Swiss Federal Institute of Technology (ETH), Zurich.
A magnesium ion-selective liquid membrane electrode based on a synthetic neutral carrier is presented. The selectivity for Mg2+ over Na+, K+, and H+ is sufficient for assays in the physiological range, but a chemometric correction of approximately 10% is still necessary for the Ca2+ interference. We optimized the membrane composition especially with respect to selectivity and accuracy by performing a new characterization procedure in which we simulated the variety of blood serum samples with 10 aqueous solutions. After a three-point calibration, these solutions were measured alternately with the middlerange calibrator for 3 min. The magnesium activity was evaluated after correcting for the calcium interference according to the Nikolsky equation. The selectivity coefficient, KpotMgCa, was itself fitted to give the minimum deviations from the assigned magnesium activities. Electrodes with an optimized membrane composition show an average deviation from theoretical activity values of about 1.6% with logKpotMgCa = -0.8.
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |