Clinical Chemistry Link to Randox Laboratories Web Site
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 39: 1831-1836, 1993;
This Article
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Parvy, P.
Right arrow Articles by Kamoun, P.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Parvy, P.
Right arrow Articles by Kamoun, P.

Clinical Chemistry, Vol 39, 1831-1836, Copyright © 1993 by American Association for Clinical Chemistry

Intra- and interlaboratory quality control for assay of amino acids in biological fluids: 14 years of the French experience

P Parvy, J Bardet, D Rabier, M Gasquet and P Kamoun
Laboratoire de Biochimie Medicale B, Hopital Necker-Enfants Malades, Paris, France.

The functioning of an external quality-control scheme for amino acids set up in 1978 is described. Two measurements were made each month by participating laboratories on a control plasma sample provided by the quality-control center; freeze-dried samples were used from 1978 to 1989 and liquid samples since 1990. In addition, two "blind" samples were sent to the laboratories each year. Every 3 months, the overall results and those of the individual laboratories were analyzed statistically. The validity of the liquid sample control is demonstrated. The progressive improvement of results is commendable. In 1990, the coefficients of variation for all participants ranged from 8.4% for alanine to 23.5% for methionine. The standards used for calibration could contribute to the broad range of results, especially those for histidine and ornithine. The use of blind samples made it possible to detect problems of calibration, of linearity of measurement, of contamination, and of identification of unusual amino acids.


The following articles in journals at HighWire Press have cited this article:


Home page
Clin. Chem.Home page
R. Venta
Year-Long Validation Study and Reference Values for Urinary Amino Acids Using a Reversed-Phase HPLC Method
Clin. Chem., March 1, 2001; 47(3): 575 - 583.
[Abstract] [Full Text] [PDF]


Home page
Nephrol Dial TransplantHome page
C. Chazot, G. Laurent, B. Charra, C. Blanc, C. VoVan, G. Jean, T. Vanel, J. C. Terrat, and M. Ruffet
Malnutrition in long-term haemodialysis survivors
Nephrol. Dial. Transplant., January 1, 2001; 16(1): 61 - 69.