Clinical Chemistry AACC Online Job Center
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 39: 1890-1893, 1993;
This Article
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Jacobs, E.
Right arrow Articles by Colman, N.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Jacobs, E.
Right arrow Articles by Colman, N.

Clinical Chemistry, Vol 39, 1890-1893, Copyright © 1993 by American Association for Clinical Chemistry

Performance of Gem Premier blood gas/electrolyte analyzer evaluated

E Jacobs, M Nowakowski and N Colman
Department of Pathology, Mount Sinai School of Medicine, Mount Sinai Hospital, New York, NY 10029.

We evaluated a new analyzer designed for point-of-care testing of blood gases, sodium, potassium, ionized calcium, and hematocrit. The Gem Premier (Mallinckrodt) system has two components: the analyzer and a disposable cartridge. Analysis takes place in the cartridge, which contains the electrochemical sensors, the calibrants, the reagents, the sampling stylus, and the waste container. The system was evaluated for imprecision and accuracy. With aqueous control materials, total imprecision (CV) was: pH, 0.10-0.18%; PCO2, 3.16-5.78%; PO2, 2.92- 4.85%; sodium, 0.82-1.44%; potassium, 1.35-1.48%; ionized calcium, 0.75- 1.45%; and hematocrit, 1.13-1.83%. Accuracy of the system was assessed by split-sample comparison with the Radiometer ABL 330 blood gas analyzer for pH and blood gases, the Nova Stat Profile 5 for whole- blood electrolyte and hematocrit analysis, and the IL Phoenix for plasma electrolyte analysis. After outlier correction, regression statistics were excellent for all analytes except sodium, which demonstrated Sy[x values between 1.80 and 2.30 mmol/L and 0.85 < or = r < 0.90.


The following articles in journals at HighWire Press have cited this article:


Home page
Clin. Chem.Home page
J. Lindemans, P. Hoefkens, A. L. van Kessel, M. Bonnay, W. R. Kulpmann, and J. D.E. van Suijlen
Portable Blood Gas and Electrolyte Analyzer Evaluated in a Multiinstitutional Study
Clin. Chem., January 1, 1999; 45(1): 111 - 117.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1993 by the American Association for Clinical Chemistry.