Clinical Chemistry
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 40: 1512-1516, 1994;
This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Linder, M. W.
Right arrow Articles by Valdes, R.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Linder, M. W.
Right arrow Articles by Valdes, R., Jr

Clinical Chemistry, Vol 40, 1512-1516, Copyright © 1994 by American Association for Clinical Chemistry

Mechanism and elimination of aspirin-induced interference in Emit II d.a.u. assays

MW Linder and R Valdes Jr
Department of Pathology, University of Louisville School of Medicine, University of Louisville Hospital, KY 40292.

The presence of salicylates in urine reduces the signal in Emit assays (Syva), potentially yielding false-negative drugs-of-abuse screening results. We demonstrate that the principal urinary metabolite of salicylate, salicyluric acid (SUA; 2-hydroxybenzoylaminoacetic acid), interferes with the measurement of NADH formed in the assay by reducing the molar absorptivity of NADH at 340 nm. Thus, for a given concentration of d.a.u. analyte the change in absorbance over the assay time interval is less in the presence of SUA. With the Emit cocaine assay on the Hitachi 704 analyzer, the rate of absorbance change (delta AR) monitored at 340 nm for a specimen containing approximately 270 micrograms/L benzoylecgonine (BE) was 57 +/- 1.9 mA/min without SUA and 29 +/- 2.7 mA/min with 5 g/L SUA (n = 20). In contrast, delta AR determined at 376 nm was 18.6 +/- 0.5 mA/min with and 17.9 +/- 0.8 mA/min without 5 g/L SUA (n = 20). Measuring the Emit assay signal at wavelengths where SUA has no absorbance (376 nm) eliminates the interference due to SUA while maintaining the precision of the assay near the cutoff concentration for BE (300 micrograms/L).


The following articles in journals at HighWire Press have cited this article:


Home page
Annals of Clinical & Laboratory ScienceHome page
G. M. Reisfield, E. Salazar, and R. L. Bertholf
Rational Use and Interpretation of Urine Drug Testing in Chronic Opioid Therapy
Ann. Clin. Lab. Sci., January 1, 2007; 37(4): 301 - 314.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1994 by the American Association for Clinical Chemistry.