Clinical Chemistry
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 40: 1805-1809, 1994;
This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Arquint, P.
Right arrow Articles by de Rooij, N. F.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Arquint, P.
Right arrow Articles by de Rooij, N. F.

Clinical Chemistry, Vol 40, 1805-1809, Copyright © 1994 by American Association for Clinical Chemistry

Micromachined analyzers on a silicon chip

P Arquint, M Koudelka-Hep, BH van der Schoot, P van der Wal and NF de Rooij
Institute of Microtechnology, University of Neuchatel, Switzerland.

For an example of a silicon-based micromachined analyzer, we describe a combined PO2, PCO2, and pH sensor designed for extracorporeal blood gas monitoring. The clinically well-accepted amperometric (PO2) and potentiometric (PCO2, pH) sensing principles are used, realized in a planar and miniaturized form on a single silicon chip (6 x 22 mm). The transducer part of the chip is fabricated by standard silicon technology. Polyacrylamide and polysiloxane polymeric layers, which are used as internal electrolyte and gas-permeable membrane, respectively, are deposited and patterned by photopolymerization. The entire sensor is fabricated on the wafer level by using integrated-circuit-compatible processes, thus allowing mass production. By integrating a flow-through channel directly on the chip, the sample size and the reagent consumption are substantially reduced. The device was characterized in aqueous solutions and in blood intended for transfusion. The sensor has a typical sensitivity of 0.36 nA/mmHg (PO2), -39 mV/decade (PCO2), and 51 mV/pH (pH); low drift; and a functional lifetime of > 2 months. The analytical precision in the physiologically expected range is better than 2 mmHg for the PO2 and PCO2 sensor, and 0.02 pH unit for the pH sensor.


The following articles in journals at HighWire Press have cited this article:


Home page
JPEN J Parenter Enteral NutrHome page
V. R. Young and A. M. Ajami
1999 Jonathan E. Rhoads lecture. Isotopic Metaprobes, Nutrition, and the Roads Ahead
JPEN J Parenter Enteral Nutr, July 1, 1999; 23(4): 175 - 194.
[Abstract] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1994 by the American Association for Clinical Chemistry.