Clinical Chemistry
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 41: 995-1003, 1995;
This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Thienpont, L. M.
Right arrow Articles by De Leenheer, A. P.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Thienpont, L. M.
Right arrow Articles by De Leenheer, A. P.

Clinical Chemistry, Vol 41, 995-1003, Copyright © 1995 by American Association for Clinical Chemistry

Candidate reference method for determining serum creatinine by isocratic HPLC: validation with isotope dilution gas chromatography- mass spectrometry and application for accuracy assessment of routine test kits

LM Thienpont, KG Van Landuyt, D Stockl and AP De Leenheer
Laboratoria voor Medische Biochemie en voor Klinische Analyse, University of Ghent, Belgium.

We present a candidate Reference Method for determining creatinine in serum, based on isocratic HPLC. The chromatographic column, with alkaline-treated aluminum oxide as stationary phase, is eluted with an equivolume mixture of methanol/acetonitrile containing 70 mL/L aqueous NaOH (10 mmol/L), and ultraviolet absorbance is detected at 240 nm. We investigated the ruggedness of the method and validated its performance with isotope dilution gas chromatography-mass spectrometry (ID GC-MS) for a set of 22 patients' sera and 10 commercially available lyophilized control materials. The mean deviation from ID GC-MS was +0.1% (range, -2.2% to +2.2%). The between-day CV, calculated from six independent measurements performed on three different days, was 0.9% (range 0.2% to 1.9%); the within-run CV was 0.8%. The total error of the method was < 3%. These performance characteristics make the method suitable for target-setting of quality-control materials and accuracy assessment of routine test kits. For the latter application, done with split-sample measurements of a panel of 83 patients' specimens, we used the Boehringer Mannheim enzymatic creatinine PAP test on the Hitachi 911, the Kodak Ektachem single-slide enzymatic creatinine test on the Ektachem 700, the Merck Jaffe kinetic assay on the Mega analyzer, and the Roche Jaffe kinetic test on the Cobas Mira.


The following articles in journals at HighWire Press have cited this article:


Home page
Clin. Chem.Home page
P. H. Petersen, D. Stockl, O. Blaabjerg, B. Pedersen, E. Birkemose, L. Thienpont, J. F. Lassen, and J. Kjeldsen
Graphical interpretation of analytical data from comparison of a field method with a Reference Method by use of difference plots
Clin. Chem., November 1, 1997; 43(11): 2039 - 2046.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1995 by the American Association for Clinical Chemistry.