Clinical Chemistry
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 42: 1659-1665, 1996;
This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Lianidou, E. S.
Right arrow Articles by Ioannou, P. C.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Lianidou, E. S.
Right arrow Articles by Ioannou, P. C.

Clinical Chemistry, Vol 42, 1659-1665, Copyright © 1996 by American Association for Clinical Chemistry

Simple spectrofluorometric determination of p-aminobenzoic and p- aminosalicylic acids in biological fluids by use of terbium-sensitized luminescence

ES Lianidou and PC Ioannou
Department of Chemistry, University of Athens, Greece.

A novel, sensitive, and selective method has been developed for determination of p-aminobenzoic (PABA) and p-aminosalicylic (PAS) acids in the N-benzoyl-L-tyrosyl-PABA/ PAS test. PAS is measured as a ternary complex with terbium and EDTA (lambda(ex) = 324 nm, lambda(em) = 546 nm) in alkaline aqueous solution (pH approximately 12.6), whereas both compounds (PABA and PAS) are measured as ternary complexes with terbium and tri-n-octylphosphine oxide (lambda(ex) = 292 nm, lambda(em) = 546 nm) in weakly acidic aqueous solution (pH approximately 5.5). We inve stigated and implemented optimum conditions for formation of these complexes, yielding respective detection limits for PABA and PAS of 0.07 and 0.02 micromol/L and ranges of application of 0-10 and 0-40 micromol/L (final concentration). The method has been successfully applied to determinations of PABA and PAS in urine and, after alkaline hydrolysis, to determinations of PABA in serum that has been deproteinized with acetonitrile. Within-run imprecision of the PABA determination ranges from 0.8% to 4.2 % for urine samples and from 3.9% to 8.2% for serum samples; day-to-day imprecision varies from 3.2% to 10% for serum samples.