Clinical Chemistry AACC Online Job Center
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 42: 1202-1205, 1996;
This Article
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Kimura, S.
Right arrow Articles by Amino, N.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Kimura, S.
Right arrow Articles by Amino, N.

Clinical Chemistry, Vol 42, 1202-1205, Copyright © 1996 by American Association for Clinical Chemistry

New enzymatic assay for calcium in serum

S Kimura, S Iyama, Y Yamaguchi, S Hayashi, R Fushimi and N Amino
Central Laboratory for Clinical Investigation, Osaka University Hospital, Japan.

We established a simple and rapid kinetic assay for measurement of calcium in serum by using urea amidolyase (EC 3.5.1.45) from yeast species. The method is based on inhibition of the enzyme by calcium. In the assay, we eliminated endogenous ammonium ion by use of glutamate dehydrogenase (GLDH; EC 1.4.1.4); then in the presence of urea amidolyase, urea, ATP, bicarbonate, magnesium, and potassium ions, ammonium ion production was inversely proportional to calcium ion concentration in serum. The concentration of ammonium ion formed was determined by adding GLDH to produce NADP+ in the presence of 2- oxoglutarate and NADPH; we then monitored the change of absorbance at 340 nm. The within-run CVs of this method were 1.7-3.2% (n = 10) at 1.53-3.08 mmol/L, respectively. Day-to-day (total) CVs were 2.8-4.1%. Analytical recovery was 92-112%. The presence of other ions, ascorbic acid, reduced glutathione, bilirubin, hemoglobin, citrate, lipemic material, or human serum albumin did not affect this assay system. The correlation between values obtained with our method (y) and o- cresolphthalein complexone method (CPC) (x) was: y = 1.001x + 0.077 mmol/L (r = 0.949, Sy[symbol: see text]x = 0.079, n = 100); with the other enzymatic method (x) it was: y = 0.952x + 0.021 mmol/L (r = 0.955, Sy[symbol: see text]x = 0.074, n = 100). The SEs for each method were: 0.025 mmol/L, our method; 0.023 mmol/L, CPC method; and 0.025 mmol/L, the other enzymatic method.


The following articles in journals at HighWire Press have cited this article:


Home page
Clin. Chem.Home page
M. Sugano, K. Yamauchi, K. Sugano, K. Kawasaki, M. Tozuka, T. Katsuyama, H. Soya, T. Tanaka, S. Imamura, and S. Nomoto
New Enzymatic Assay Using Phospholipase D to Measure Total Calcium in Serum
Clin. Chem., June 1, 2005; 51(6): 1021 - 1024.
[Full Text] [PDF]


Home page
Clin. Chem.Home page
Y. Morishita, Y. Iinuma, N. Nakashima, A. Kadota, A. Miike, and T. Tadano
Enzymatic Assay of Calcium in Serum with Phospholipase D
Clin. Chem., December 1, 1999; 45(12): 2280 - 2283.
[Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1996 by the American Association for Clinical Chemistry.