Clinical Chemistry Link to Randox Laboratories Web Site
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 43: 2064-2075, 1997;
This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via ISI Web of Science (17)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Dahler-Eriksen, B. S.
Right arrow Articles by Brandslund, I.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Dahler-Eriksen, B. S.
Right arrow Articles by Brandslund, I.
Related Collections
Right arrow Laboratory Management
Right arrow Evidence Based Laboratory Medicine and Test Utilization
Right arrow Proteomics and Protein Markers
(Clinical Chemistry. 1997;43:2064-2075.)
© 1997 American Association for Clinical Chemistry, Inc.


Articles

Evaluation of a near-patient test for C-reactive protein used in daily routine in primary healthcare by use of difference plots

Bjarne Steen Dahler-Eriksen1,2,a, Jens Flensted Lassen1, Per Hyltoft Petersen3, Erik Dalsgaard Lund1, Torsten Lauritzen2 and Ivan Brandslund1

1 Department of Clinical Chemistry, Vejle County Central Hospital, DK-7100 Vejle, Denmark.

2 Department of General Practice, University of Aarhus, Aarhus, Denmark.

3 Department of Clinical Chemistry, Odense University Hospital, Denmark.
a Author for correspondence. Fax + 45 75 82 18 14.

We have assessed the technical performance and robustness of NycoCard® CRP Whole Blood, a near-patient test for C-reactive protein (CRP), when used in realistic daily routine situations in general practice clinics (GPC). Thirteen GPCs participated, five of them with technician staff. From 898 patients, split-sample measurements for CRP were made. Results from GPCs were compared with results from a turbidimetric laboratory method, traceable to international reference preparations (IFCC CRM 470). Results were evaluated in difference plots where the expected distribution, due to an estimated analytical variation, was compared with measured differences. Of all difference points, 91.5% (n = 819) were within a 95% prediction interval based on the imprecision of both methods. Mean bias (95% confidence interval) was -0.3 mg/L (-0.9 to 0.3). No differences in analytic quality were found between GPCs with technician staffs and GPCs without, and between test results obtained within the first and second week, compared with the rest of the study period. We find the test as good when used in GPCs as could be expected from laboratory testing, and consequently robust, which is a necessity for use in routine situations in general practice. General application of difference plots in test evaluations are discussed in detail.




The following articles in journals at HighWire Press have cited this article:


Home page
Clin. Chem.Home page
S. Skeie, G. Thue, K. Nerhus, and S. Sandberg
Instruments for Self-Monitoring of Blood Glucose: Comparisons of Testing Quality Achieved by Patients and a Technician
Clin. Chem., July 1, 2002; 48(7): 994 - 1003.
[Abstract] [Full Text] [PDF]


Home page
Clin. Chem.Home page
G. Soletormos, P. Hyltoft Petersen, and P. Dombernowsky
Progression Criteria for Cancer Antigen 15.3 and Carcinoembryonic Antigen in Metastatic Breast Cancer Compared by Computer Simulation of Marker Data
Clin. Chem., July 1, 2000; 46(7): 939 - 949.
[Abstract] [Full Text] [PDF]


Home page
Clin. Chem.Home page
B. S. Dahler-Eriksen, T. Lauritzen, J. F. Lassen, E. D. Lund, and I. Brandslund
Near-Patient Test for C-Reactive Protein in General Practice: Assessment of Clinical, Organizational, and Economic Outcomes
Clin. Chem., April 1, 1999; 45(4): 478 - 485.
[Abstract] [Full Text] [PDF]


Home page
Clin. Chem.Home page
M. Andersen, P. H. Petersen, O. Blaabjerg, J. Hangaard, and C. Hagen
Evaluation of growth hormone assays using ratio plots
Clin. Chem., May 1, 1998; 44(5): 1032 - 1038.
[Abstract] [Full Text] [PDF]


Home page
Clin. Chem.Home page
R. Gambino
C-Reactive Protein—Undervalued, Underutilized
Clin. Chem., November 1, 1997; 43(11): 2017 - 2018.
[Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1997 by the American Association for Clinical Chemistry.