Articles |
Athletic Drug Testing and Toxicology Laboratory, Indiana University Medical Center, Medical Science Bldg. A-128, 635 Barnhill Dr., Indianapolis, IN 46202-5120.
a Author for correspondence. Fax 317-274-3223; e-mail lbowers{at}iupui.edu
Limits of detection are important issues for GC/MS screening for anabolic agents and for confirmation of various drugs of abuse. We compared a quadrupole ion trap (QIT) operated in two different selected-ion storage modes and a quadrupole mass filter (QMF) operated in the selected-ion monitoring mode. Results with the model compound tetrachlorobenzene indicate that, for simultaneous monitoring of more than four ions, the QIT operated in a frequency-modulated selected-ion storage mode has better limits of detection than the QMF. Use of a single-ion storage technique gave results similar to those of the QMF. We also evaluated both QIT selected-ion storage approaches for the limits of detection of the trimethylsilyl derivatives of four anabolic steroid metabolites and the ß-agonist clenbuterol. We found no improvement in detection limits over that of a similar method with selected-ion monitoring and a QMF when four anabolic steroid metabolites and clenbuterol were extracted from a urine matrix. The lack of improvement in the limit of detection resulted from matrix background signals at masses similar to those of the steroids.
Key Words: indexing terms: quadrupole ion trap quadrupole mass filter GC-MS abused drugs clenbuterol
The following articles in journals at HighWire Press have cited this article:
![]() |
L. D. Bowers Analytical advances in detection of performance-enhancing compounds Clin. Chem., July 1, 1997; 43(7): 1299 - 1304. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. L. Fitzgerald, D. A. Herold, and A. L. Yergey Trade-offs in Mass Spectrometry Clin. Chem., June 1, 1997; 43(6): 915 - 915. [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |