Clinical Chemistry
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 45: 1557-1561, 1999;
This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Guaita, S.
Right arrow Articles by Camps, J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Guaita, S.
Right arrow Articles by Camps, J.
Related Collections
Right arrow Clinical Immunology
Right arrow Proteomics and Protein Markers
Right arrow Automation and Analytical Techniques
(Clinical Chemistry. 1999;45:1557-1561.)
© 1999 American Association for Clinical Chemistry, Inc.


Articles

Evaluation of a Particle-enhanced Turbidimetric Immunoassay for the Measurement of Immunoglobulin E in an ILab 900 Analyzer

Sandra Guaita, Josep MA Simó, Natàlia Ferré, Jorge Joven and Jordi Campsa

Centre de Recerca Biomèdica, Hospital Universitari de Sant Joan, Carrer Sant Joan s/n, 43201-Reus, Catalunya, Spain.
a Author for correspondence. Fax 34-77-31-25-69;

Background: The measurement of immunoglobulin E (IgE) in serum is widely used in the diagnosis of allergic reactions and parasitic infections. We describe here a fully automated assay for human IgE suitable for routine application in a general chemistry analyzer.

Methods: We used an ILab 900® analyzer. This instrument automates a particle-enhanced immunoturbidimetric assay with an analysis time of 9 min.

Results: The assay was linear in the range 4–1000 kIU/L (r = 0.9998). The intra- and interassay CVs at 57, 235, and 434 kIU/L were <3.5% and <7.4%, respectively. The detection limit was 4 kIU/L. Hemoglobin (<=16 g/L), bilirubin (<=250 µmol/L), and myeloma paraproteins did not interfere with the assay. The assay showed good correlation with a microparticle enzyme immunoassay (r = 0.998) with a mean difference between methods of -6 ± 26 kIU/L.

Conclusion: The new automated serum assay for IgE is an attractive alternative that avoids the need for dedicated instrumentation.







HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS