Articles |
Methods: Theoretical equations based on the Deming approach, further developed by physicists and extended herein, were applied to method-comparison data analysis. Monte Carlo simulations were used to demonstrate the validity of the new procedure and to compare its performance to ordinary linear regression (OLR) and simple Deming regression (SDR) procedures.
Results: Simulation studies included three types of data commonly encountered in method-comparison studies: (a) constant within-method SDs for both methods, (b) constant within-method CVs for both methods, and (c) neither SDs nor CVs constant for both methods. For all cases examined, OLR produced unreliable confidence intervals of the estimated bias. However, OLR point estimates of systematic bias were reliable when the correlation coefficient was >0.975. SDR produced reliable estimates of systematic bias for all cases studied, but the confidence intervals of systematic bias were unreliable when SDs of methods varied as a function of analyte concentration.
Conclusion: Only iteratively reweighted general Deming regression produced statistically unbiased estimates of systematic bias and reliable confidence intervals of bias for all cases.
The following articles in journals at HighWire Press have cited this article:
![]() |
S.-T. Lee, C. W. Weykamp, Y.-W. Lee, J.-W. Kim, and C.-S. Ki Effects of 7 Hemoglobin Variants on the Measurement of Glycohemoglobin by 14 Analytical Methods Clin. Chem., December 1, 2007; 53(12): 2202 - 2205. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. UMETANI, S. HIRAMATSU, and D. S.B HOON Higher Amount of Free Circulating DNA in Serum than in Plasma Is Not Mainly Caused by Contaminated Extraneous DNA during Separation. Ann. N.Y. Acad. Sci., September 1, 2006; 1075: 299 - 307. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Ferre, J. Camps, J. Marsillach, B. Mackness, M. Mackness, B. Coll, M. Tous, and J. Joven Comparison of Paraoxonase 1 Measurements in Serum and in Lithium-Heparin-Anticoagulated Plasma Samples Clin. Chem., May 1, 2005; 51(5): 922 - 923. [Full Text] [PDF] |
||||
![]() |
S. A. Santini, C. Carrozza, C. Vulpio, E. Capoluongo, G. Luciani, P. Lulli, B. Giardina, and C. Zuppi Assessment of Parathyroid Function in Clinical Practice: Which Parathyroid Hormone Assay Is Better? Clin. Chem., July 1, 2004; 50(7): 1247 - 1250. [Full Text] [PDF] |
||||
![]() |
P. Suominen Evaluation of an Enzyme Immunometric Assay to Measure Serum Adiponectin Concentrations Clin. Chem., January 1, 2004; 50(1): 219 - 221. [Full Text] [PDF] |
||||
![]() |
C. Gay-Bellile, D. Bengoufa, P. Houze, D. Le Carrer, M. Benlakehal, B. Bousquet, B. Gourmel, and T. Le Bricon Automated Multicapillary Electrophoresis for Analysis of Human Serum Proteins Clin. Chem., November 1, 2003; 49(11): 1909 - 1915. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Khuseyinova, A. Imhof, G. Trischler, D. Rothenbacher, W. L. Hutchinson, M. B. Pepys, and W. Koenig Determination of C-Reactive Protein: Comparison of Three High-Sensitivity Immunoassays Clin. Chem., October 1, 2003; 49(10): 1691 - 1695. [Full Text] [PDF] |
||||
![]() |
C. Bieglmayer, G. Prager, and B. Niederle Kinetic Analyses of Parathyroid Hormone Clearance as Measured by Three Rapid Immunoassays during Parathyroidectomy Clin. Chem., October 1, 2002; 48(10): 1731 - 1738. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. M. Flanders, R. Crist, S. Safapour, and G. M. Rodgers Evaluation and Performance Characteristics of the STA-R Coagulation Analyzer Clin. Chem., September 1, 2002; 48(9): 1622 - 1624. [Full Text] [PDF] |
||||
![]() |
S. Rothkrantz-Kos, M. P.J. Schmitz, O. Bekers, P. P.C.A. Menheere, and M. P. van Dieijen-Visser High-Sensitivity C-Reactive Protein Methods Examined Clin. Chem., February 1, 2002; 48(2): 359 - 362. [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |