Background: The total protein content of urine is a good index of renal function, but its determination is unreliable. Protein dye-binding assays are simple, but they characteristically lack a uniform response to different proteins.
Methods: We investigated a differential response of the Sigma Microprotein Coomassie Brilliant Blue (CBB) and Pyrogallol Red-molybdate (PRM) protein dye-binding assays to urine, using human albumin, albumin/globulin, or urinary protein as calibrator.
Results: The urine protein values (n = 60) obtained with the CBB assay were 11013 500 mg/L (mean, 2390 mg/L) compared with 16018 300 mg/L (mean, 3470 mg/L) obtained with the PRM assay (CBB:PRM protein concentration ratio, 0.460.88, mean, 0.69 ± 0.10). The differential response was highly reproducible as indicated by Sigma urine control Level 1 (within-day CBB:PRM ratio, 0.68 ± 0.02; between-day CBB:PRM ratio, 0.67 ± 0.04) and Sigma urine control Level 2 (within-day CBB:PRM ratio, 0.60 ± 0.01; between-day CBB:PRM ratio, 0.59 ± 0.02). The use of urinary protein as a calibrator (rather than human albumin) greatly improved the agreement between the assays when applied to urine (yCBB = 0.972xPRM - 16 vs yCBB = 0.685xPRM + 17). In studies using urine controls, this calibrator also improved agreement between the CBB, PRM, trichloroacetic acid (TCA), and benzethonium chloride protein methods and, to a lesser extent, agreement with the TCA-Ponceau S method.
Conclusion: The use of a urinary protein calibrator improves the agreement between different methods used to determine total protein in urine.
The following articles in journals at HighWire Press have cited this article:
![]() |
A. Leanos-Miranda, J. Marquez-Acosta, F. Romero-Arauz, G. M. Cardenas-Mondragon, R. Rivera-Leanos, I. Isordia-Salas, and A. Ulloa-Aguirre Protein:Creatinine Ratio in Random Urine Samples Is a Reliable Marker of Increased 24-Hour Protein Excretion in Hospitalized Women with Hypertensive Disorders of Pregnancy Clin. Chem., September 1, 2007; 53(9): 1623 - 1628. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Maachi, S. Fellahi, A. Regeniter, M.-E. Diop, J. Capeau, J. Rossert, and J.-P. Bastard Patterns of Proteinuria: Urinary Sodium Dodecyl Sulfate Electrophoresis Versus Immunonephelometric Protein Marker Measurement Followed by Interpretation with the Knowledge-Based System MDI-LabLink Clin. Chem., October 1, 2004; 50(10): 1834 - 1837. [Full Text] [PDF] |
||||
![]() |
F. M. Yilmaz, N. Celebi, and D. Yucel Automated Turbidimetric Benzalkonium Chloride Method for Measurement of Protein in Urine and Cerebrospinal Fluid Clin. Chem., August 1, 2004; 50(8): 1450 - 1452. [Full Text] [PDF] |
||||
![]() |
R. B. Caldwell and C. T. Lattemann Simple and Reliable Method To Precipitate Proteins from Bacterial Culture Supernatant Appl. Envir. Microbiol., January 1, 2004; 70(1): 610 - 612. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Marshall and K. M. Williams Total Protein Determination in Urine: Aminoglycoside Interference Clin. Chem., January 1, 2003; 49(1): 202 - 203. [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |